| Research article summary (published 30 Mar 2006): |
Evaluation of electromagnetic interference of GSM mobile phones with pacemakers featuring remote monitoring functions.
Full Abstract
The aim of this study was to evaluate whether global system for mobile communication (GSM) cellular phones can affect the home monitoring (HM) function implemented in last generation pacemakers (PM). Methods:
We performed in vitro and in vivo tests on the BA03 PM (Biotronik GmbH, Berlin, Germany). In vitro we evaluated whether an improper HM procedure or an altered patient-activated HM function occurred. We used two cellular phone models, with a fixed external or internal antenna, positioned close to both the PM and the mobile phone-like device, during handover, ringing, and talking. All the tests were done with the PM in air, at 900 and 1,800 MHz GSM bands, under worst case conditions. A subset of these tests was repeated in 17 patients:
the mobile phones were moved both around the PM implant site and the mobile cell phone-like device, during talking. Results:In vitro, neither the HM procedure nor PM functioning were corrupted by the GSM communications:
all the transmissions were correctly received, with a maximum transmission delay of about 110 seconds. In vivo, the rate of successful transmissions was 93%. Conclusion:
Our data show that HM function does not call for specific restrictions on the use of GSM cellular phones.
Learn Faster Today Improve your study skills
Author information
Author/s: Calcagnini, G (G); Censi, F (F); Floris, M (M); Pignalberi, C (C); Ricci, R (R); Biancalana, G (G); Bartolini, P (P); Santini, M (M);
Affiliation: Department of Technologies and Health--Istituto Superiore Sanità, Viale Regina Elena 299, 00161, Roma, Italy. giovanni.calcagnini(-atsign-)iss.it
Journal and publication information
Publication Type: Evaluation Studies; Journal Article; Research Support, Non-U.S. Gov't
Journal: Pacing and clinical electrophysiology : PACE (Pacing Clin Electrophysiol), published in United States. (Language: eng)
Reference: 2006-Apr; vol 29 (issue 4) : pp 380-5
Dates: Created 2006/05/02; Completed 2006/09/05; Revised 2006/11/15;
PMID: 16650266, status: MEDLINE (last retrieval date: 12/26/2008)
Sourced from the National Library of Medicine. Abstract text and other information may be subject to copyright.
External Links for this article (including full text providers, if available):
Click Electronic Full-text Provider Links to see options for finding the electronic full text links to this article. Note there may be a subscription or fee required for access to the full text. See our FAQ for information on finding FREE full text articles.
This article may also be located in paper journal collections available in many libraries. Use the Journal and Publication Information above to find the full article.
MeSH headings (categories)
This article was linked to the MESH Headings shown below.
|
|
Related articles
These are the highest related articles currently in the database:
- The fantastic pacemaker system.
30 Aug 2002 - Implementation of a WAP-based telemedicine system for patient monitoring.
30 May 2003 - [Home Monitoring with implantable ICD--a diagnostic innovation?]
30 Aug 2005 - New horizons in ambulatory electroencephalography.
29 Apr 2003 - [The Herz Handy--a new telemedical service concept for heart patients]
30 Aug 2005 - Wireless Holter transmission in suspected dysrhythmias.
15 Aug 2006 - Design of wearable cardiac telemedicine system.
30 Dec 2006 - A new mobile phone-based ECG monitoring system.
30 Dec 2006 - [The technical possibilities in telemonitoring of physiological parameters]
30 Aug 2005 - A mobile phone-based ECG monitoring system.
30 Dec 2005
Related Article Map
Legend:
- FREE Full text Article.
- Abstract only.
- Title only. More help.
See a large map of 100+ related articles.