PURPOSE:
This study aimed to demonstrate how the level of accuracy in intraocular lens (IOL) power calculation can be improved with optical biometry using partial optical coherence interferometry (PCI) (Zeiss IOLMaster) and current anterior chamber depth (ACD) prediction algorithms.

METHODS:
Intraocular lens power in 461 consecutive cataract operations was calculated using both PCI and ultrasound and the accuracy of the results of each technique were compared. To illustrate the importance of ACD prediction per se, predictions were calculated using both a recently published 5-variable method and the Haigis 2-variable method and the results compared. All calculations were optimized in retrospect to account for systematic errors, including IOL constants and other off-set errors.

RESULTS:
The average absolute IOL prediction error (observed minus expected refraction) was 0.65 dioptres with ultrasound and 0.43 D with PCI using the 5-variable ACD prediction method (p < 0.00001). The number of predictions within +/- 0.5 D, +/- 1.0 D and +/- 2.0 D of the expected outcome was 62.5%, 92.4% and 99.9% with PCI, compared with 45.5%, 77.3% and 98.4% with ultrasound, respectively (p < 0.00001). The 2-variable ACD method resulted in an average error in PCI predictions of 0.46 D, which was significantly higher than the error in the 5-variable method (p < 0.001).

CONCLUSIONS:
The accuracy of IOL power calculation can be significantly improved using calibrated axial length readings obtained with PCI and modern IOL power calculation formulas incorporating the latest generation ACD prediction algorithms.

 

Learn Faster Today      Improve your study skills

Author information

Author/s: Olsen, Thomas (T);

Affiliation: University Eye Clinic, Aarhus Kommunehospital, Aarhus, Denmark. tkolsen(-atsign-)dadlnet.dk

Journal and publication information

Publication Type: Journal Article; Research Support, Non-U.S. Gov't

Journal: Acta ophthalmologica Scandinavica (Acta Ophthalmol Scand), published in Denmark. (Language: eng)

Reference: 2007-Feb; vol 85 (issue 1) : pp 84-7

Dates: Created 2007/01/24; Completed 2007/03/15; Revised 2008/11/21;

PMID: 17244216, status: MEDLINE (last retrieval date: 12/26/2008)

Sourced from the National Library of Medicine. Abstract text and other information may be subject to copyright.

Comments and Corrections

CommentIn: Acta Ophthalmol Scand. 2007 Nov;85(7):803-4; author reply 804-5. (PMID: 17822447)

External Links for this article (including full text providers, if available):

Click Electronic Full-text Provider Links to see options for finding the electronic full text links to this article. Note there may be a subscription or fee required for access to the full text. See our FAQ for information on finding FREE full text articles.

This article may also be located in paper journal collections available in many libraries. Use the Journal and Publication Information above to find the full article.

MeSH headings (categories)

This article was linked to the MESH Headings shown below.

Related articles

These are the highest related articles currently in the database:

See 100+ related articles.

Related Article Map

11/29/2005
11/6/2007
Higher Relevance Score (14)
Lower Relevance Score (10)

Legend: - FREE Full text Article. - Abstract only. - Title only. More help.

See a large map of 100+ related articles.

© Advanogy.com 2003-2009 (ACN 104 198 263) - All rights reserved. Terms of Use | Contact Us | Index