| Research article summary (published 29 Nov 2006): |
The role of post-implant dosimetry in the quality assessment of prostate implants. The RAH experience.
Full Abstract
The quality of a prostate implant using radioactive I-125 seeds can be assessed by performing post-implant dosimetry, which therefore has to be a reliable evaluation tool. At the RAH, we have noticed large variations in post-implant dosimetric parameters compared to the pre-implant data. The purpose of this study was to investigate the cause of these differences. Post-implant dosimetry was performed 4 weeks after implant on 15 prostate cancer patients. The CT images of the pelvis were exported to the planning system, where contouring of the target (prostate) was executed by 6 clinicians, followed by the post-implant dosimetry performed by a physicist. Pre- and post-implant dosimetric parameters were analyzed and compared. The average target volume based on the ultrasound measurements was 35.8 cc. Post-implant CT volumes were determined and averaged over the 15 patients by each clinician, and their average values vary from 28.9 cc to 67.9 cc. Beside the under/overestimation of the target on the CT there was also a "shift" in the target base on the ultrasound image. By comparing pre-implant and post-implant dosimetric parameters we have encountered a significant discrepancy between target volumes based on ultrasound image and CT image. It was concluded that the accuracy of target coverage was partially connected to the poorer quality of the CT image compared to the ultrasound scan, patient's anatomy, but mostly to the poor implantation of the base.
Learn Faster Today Improve your study skills
Author information
Author/s: Marcu, L (L); Quach, K (K);
Affiliation: University of Adelaide, School of Chemistry and Physics, Adelaide, Australia. lmarcu(-atsign-)mail.rah.sa.gov.au
Journal and publication information
Publication Type: Journal Article
Journal: Australasian physical & engineering sciences in medicine / supported by the Australasian College of Physical Scientists in Medicine and the Australasian Association of Physical Sciences in Medicine (Australas Phys Eng Sci Med), published in Australia. (Language: eng)
Reference: 2006-Dec; vol 29 (issue 4) : pp 310-4
Dates: Created 2007/01/30; Completed 2007/02/22;
PMID: 17260585, status: MEDLINE (last retrieval date: 12/26/2008)
Sourced from the National Library of Medicine. Abstract text and other information may be subject to copyright.
External Links for this article (including full text providers, if available):
Click Electronic Full-text Provider Links to see options for finding the electronic full text links to this article. Note there may be a subscription or fee required for access to the full text. See our FAQ for information on finding FREE full text articles.
This article may also be located in paper journal collections available in many libraries. Use the Journal and Publication Information above to find the full article.
MeSH headings (categories)
This article was linked to the MESH Headings shown below.
|
|
Related articles
These are the highest related articles currently in the database:
- Monte Carlo calculated TG-43 dosimetry parameters for the SeedLink 125Iodine brachytherapy system.
30 Aug 2003 - Calibration procedures for seeds preloaded in cartridges.
30 Jul 2006 - A semi-analytic approach to determine dose rate constant of brachytherapy sources in compliance with AAPM TG 60 formalism.
29 Nov 2003 - Calibration of the NPL secondary standard radionuclide calibrator for 125I seeds used for prostate brachytherapy. National Physical Laboratory.
30 Dec 2001 - Improved radial dose function estimation using current version MCNP Monte-Carlo simulation: Model 6711 and ISC3500 125I brachytherapy sources.
29 Nov 2004 - On the development of consensus values of reference dosimetry parameters for interstitial brachytherapy sources.
29 Apr 2004 - Development of an Australian secondary standard for the reference air kerma rate measurement of 125I seeds.
30 Aug 2005 - Dose escalation in permanent brachytherapy for prostate cancer: dosimetric and biological considerations.
5 Sep 2003 - Calculation of 18F, 99mTc, 111In and 123I calibration factor using the penelope ionization chamber simulation method.
30 Jan 2004 - Verification of the agreement of two dosimetric methods with radioiodine therapy in hyperthyroid patients.
30 Jul 2006