BACKGROUND:
The technique of sentinel node biopsy (SNB) may be used as a staging technique for cancer patients. SNB involves the localization of lymph nodes that have accumulated a radioactive tracer, and this requires surgeons to gain and demonstrate skills in the use of hand-held radiation detectors (gamma probes).

METHODS:
A computerized gamma probe simulator (GAPS) has been developed as a training tool to provide realistic simulations of the clinical distribution of radioactivity, without the use of radioactive materials, and a realistic simulation of the count rate response from a gamma probe. Electromagnetic spatial tracking of the surgeon's movement of the probe allows objective assessment of the user's accuracy in localizing a virtual sentinel node. The physical accuracy of the simulation has been validated, with a spatial accuracy of (0.06 mm), and a count rate error of 0.28%.

RESULTS:
The GAPS has been used in the training of 94 breast surgeons, with a mean error node localization of 3.8 mm (range, 0.1-16 mm) and a mean search time of 131 s (range 36-314 s), showing that objective feedback on performance can be given by the system. Modification to train for other sentinel node applications is simple, and the system has been used for training in the application of penile sentinel node surgery.

CONCLUSION:
A computerized gamma probe simulator has been developed which provides realistic training tasks for surgeons in sentinel node biopsy, with a wide range of simulated clinical cases, allowing the objective assessment of a trainee's performance in the use of a hand-held gamma probe without the use of radioactive sources.

 

Learn Faster Today      Improve your study skills

Author information

Author/s: Britten, Alan (A); Newey, Valentine R (VR); Clarke, Ron (R);

Affiliation: Medical Physics Department, St George's Hospital, London, UK.

Journal and publication information

Publication Type: Journal Article

Journal: Nuclear medicine communications (Nucl Med Commun), published in England. (Language: eng)

Reference: 2007-Mar; vol 28 (issue 3) : pp 225-9

Dates: Created 2007/01/31; Completed 2007/04/02;

PMID: 17264782, status: MEDLINE (last retrieval date: 12/26/2008)

Sourced from the National Library of Medicine. Abstract text and other information may be subject to copyright.

External Links for this article (including full text providers, if available):

Click Electronic Full-text Provider Links to see options for finding the electronic full text links to this article. Note there may be a subscription or fee required for access to the full text. See our FAQ for information on finding FREE full text articles.

This article may also be located in paper journal collections available in many libraries. Use the Journal and Publication Information above to find the full article.

MeSH headings (categories)

This article was linked to the MESH Headings shown below.

Associated Chemicals: Radiopharmaceuticals (0)

Related articles

These are the highest related articles currently in the database:

See 100+ related articles.

Related Article Map

6/29/2000
3/30/2007
Higher Relevance Score (9)
Lower Relevance Score (5)

Legend: - FREE Full text Article. - Abstract only. - Title only. More help.

See a large map of 100+ related articles.

© Advanogy.com 2003-2009 (ACN 104 198 263) - All rights reserved. Terms of Use | Contact Us | Index