Comparison of predictions made by the intraocular lens master and ultrasound biometry.
Full Abstract
OBJECTIVE:
To ascertain whether a new instrument that uses partial coherence interferometry technology (Intraocular Lens Master [IOLm]) or ultrasound biometry provides a more accurate prediction of refractive outcomes in cataract surgery.
METHODS:
This was a retrospective medical record review of 421 eyes of 304 patients who underwent cataract surgery with the IOLm and ultrasound biometry from January 3, 2002, to December 2, 2005. The mean difference between the prediction with each technology and the final spherical equivalent was compared. Circumstances in which both technologies failed to make accurate predictions were investigated.
RESULTS:
The mean (SD) of the difference between predicted refraction and final spherical equivalent was -0.43 (0.84) diopters (D) for the IOLm and -0.60 (0.87) D for ultrasound biometry, indicating that on average the IOLm was a closer predictor than ultrasound biometry of the final spherical equivalent (P < .001). The IOLm had a 5% higher likelihood of predicting a spherical equivalent within 0.25 D than ultrasound biometry (P = .06), an 8% higher likelihood of predicting a spherical equivalent within 0.50 D (P < .001), and an 8% higher likelihood of predicting a spherical equivalent within 1.00 D (P < .001).
CONCLUSIONS:
To our knowledge, this is the most eyes examined for cataract surgery in a prospective fashion with both the IOLm and ultrasound biometry. The IOLm is a better predictor of postoperative refraction than ultrasound biometry, particularly within close ranges.
Learn Faster Today Improve your study skills
Author information
Author/s: Bhatt, Anand B (AB); Schefler, Amy C (AC); Feuer, William J (WJ); Yoo, Sonia H (SH); Murray, Timothy G (TG);
Affiliation: Bascom Palmer Eye Institute, Department of Ophthalmology, University of Miami School of Medicine, 900 NW 17th St, Miami, FL 33136, USA.
Journal and publication information
Publication Type: Comparative Study; Journal Article
Journal: Archives of ophthalmology (Arch Ophthalmol), published in United States. (Language: eng)
Reference: 2008-Jul; vol 126 (issue 7) : pp 929-33
Dates: Created 2008/07/15; Completed 2008/07/24;
PMID: 18625938, status: MEDLINE (last retrieval date: 11/6/2008)
Sourced from the National Library of Medicine. Abstract text and other information may be subject to copyright.
External Links for this article (including full text providers, if available):
Click Electronic Full-text Provider Links to see options for finding the electronic full text links to this article. Note there may be a subscription or fee required for access to the full text. See our FAQ for information on finding FREE full text articles.
This article may also be located in paper journal collections available in many libraries. Use the Journal and Publication Information above to find the full article.
MeSH headings (categories)
This article was linked to the MESH Headings shown below.
|
|
Related articles
These are the highest related articles currently in the database:
- Comparison of postoperative refractive outcome in phacotrabeculectomy and phacoemulsification with posterior chamber intraocular lens implantation.
30 Jan 2006 - Improved accuracy of intraocular lens power calculation with the Zeiss IOLMaster.
30 Jan 2007 - Biometry and intraocular lens power calculation.
30 Dec 2007 - Keratometry, biometry and prediction of intraocular lens power in the equine eye.
30 Aug 2006 - Accuracy of intraocular lens power prediction using the Hoffer Q, Holladay 1, Holladay 2, and SRK/T formulas.
29 Nov 2006 - Biometry and formula accuracy with intraocular lenses used for cataract surgery in extreme hyperopia.
30 May 2007 - Performance of conventional biometry vs. integrated laser interferometry with keratometry device in intraocular lens measurement.
30 Oct 2007 - [Optical biometry of eyes corrected by phakic intraocular lenses]
29 Nov 2005 - Optimization of biometry for intraocular lens implantation using the Zeiss IOLMaster.
30 Jul 2005 - Optimization of biometry for intraocular lens implantation using the Zeiss IOLMaster.
29 Sep 2005
Related Article Map
Legend:
- FREE Full text Article.
- Abstract only.
- Title only. More help.
See a large map of 100+ related articles.