| Research article summary (published 29 Sep 2009): |
Impact of multiaccess infusion devices on in vitro drug delivery during multi-infusion therapy.
Full Abstract
BACKGROUND: Multiaccess infusion sets allow multiple simultaneous infusions but may induce interference in drug delivery resulting from large variations in the delivery rate of potent drugs. In this study, we sought to understand the influence of multiaccess infusion device properties (dead space volume and antireflux valve [ARV]) on drug delivery during multi-infusion therapy. METHODS: Infusion sets differing in length, dead space volume, and presence of an ARV were assessed. Three drugs were infused simultaneously through different access points, and their concentrations were obtained using UV spectrophotometric analysis of the effluent. Different infusion configurations were compared by assessing (1) the amount of drug delivered to the patient per unit of time, (2) the mean amount of drug delivered to the patient per unit of time during the steady-state infusion (mass flow rate plateau), and (3) flow change efficiency calculated from the ratio of the area under the experimental instant mass flow rate curve to the area corresponding to theoretical instant mass flow rate curve. RESULTS: Infusion sets with lower dead space volumes offered significantly higher flow change efficiency (53.0% +/- 15.4% with a dead space volume equal to 0.046 mL 5 min after the start of infusion) than infusion sets with higher dead space volume (5.6% +/- 8.2% with a dead space volume equal to 6.16 mL), whatever the flow rate changes. Even in case of large dead space volumes, the presence of an ARV significantly increased the mass flow rate plateau (from 92.4% to 99.3% of the theoretical plateau without and with the presence of an ARV, respectively). CONCLUSIONS: Multi-infusion therapy induces perturbation in drug delivery. These perturbations (lag time, backflow, and bolus) could be reduced by using infusion sets including very low dead space volume and an ARV.
Author information
Author/s: Décaudin, Bertrand (B); Dewulf, Sophie (S); Lannoy, Damien (D); Simon, Nicolas (N); Secq, Alexandre (A); Barthélémy, Christine (C); Debaene, Bertrand (B); Odou, Pascal (P);
Affiliation: Department of Biopharmacy, Galenic and Hospital Pharmacy EA 4034-IFR114, Lille 2 University, Lille, France.
Journal and publication information
Publication Type: Journal Article
Journal: Anesthesia and analgesia (Anesth Analg), published in United States. (Language: eng)
Reference: 2009-Oct; vol 109 (issue 4) : pp 1147-55
Dates: Created 2009/09/18; Completed 2009/10/01;
PMID: 19762742, status: MEDLINE (last retrieval date: 10/1/2009, IMS Date: )
Sourced from the National Library of Medicine. Abstract text and other information may be subject to copyright.
External Links for this article
(including full text providers, if available):
Click Electronic Full-text Provider Links to see options for finding the electronic full text links to this article. Note there may be a subscription or fee required for access to the full text. See our FAQ for information on finding FREE full text articles.
This article may also be located in paper journal collections available in many libraries. Use the Journal and Publication Information above to find the full article.
MeSH headings (categories)
This article was linked to the MESH Headings shown below.
Related articles
These are the highest related articles currently in the database:
- Loss of glyceryl trinitrate and of isosorbide dinitrate during infusion: a literature survey and practical recommendations.
30 Dec 1982 - Accurate continuous drug delivery at low infusion rate with a novel microvolumetric infusion pump (MVIP): pump design, evaluation and comparison to the current standard.
30 Oct 2004 - Prediction of convection-enhanced drug delivery to the human brain.
12 Sep 2007 - The Panomat P-10 micro-volumetric infusion pump is suitable for continuous drug administration at minimal flow rates.
29 Nov 2002 - Administering a 24-hour supply of antibiotics with a programmable, automated syringe pump.
29 Nov 1985 - Flow rate, syringe size and architecture are critical to start-up performance of syringe pumps.
28 Jan 2007 - Protein delivery with infusion pumps.
30 Dec 1996 - Performance of selected flow-restricting infusion devices.
13 Feb 1998 - Novel approaches to intraperitoneal drug delivery.
30 Dec 1988 - [Automatic syringe for intravascular infusions]
30 Dec 1974
Related Article Map
Legend:
- FREE Full text Article.
- Abstract only.
- Title only. More help.
See a large map of 100+ related articles.