An in vitro analysis of central venous drug delivery by continuous infusion: the effect of manifold design and port selection.
Full Abstract
BACKGROUND: Central venous catheters are used extensively in anesthesia and critical care. Multiport manifolds allow for simultaneous administration of multiple medication infusions into a common central venous catheter lumen. The structures of such manifolds vary considerably. In this study, we quantitatively compared, in a laboratory model of continuous drug infusion, the drug delivery dynamics of a traditional stopcock manifold and a microinfusion manifold constructed to minimize dead volume. METHODS: A syringe pump infused a saline carrier solution at a low flow rate frequently used in an intensive care unit (10 mL/h) through a multiport manifold connected to the 16-gauge lumen of a standard 16-cm triple-lumen catheter. The model drug methylene blue (3 mL/h) joined the carrier flow at the first, second, or fourth stopcock of a traditional manifold or 1 of 2 positions in a microinfusion manifold, a new device designed to minimize dead volume. Effluent samples were collected every minute for quantitative spectrophotometric analysis of delivery onset and offset. RESULTS: Onset and offset times differed significantly among individual ports of the traditional 4-stopcock manifold. There was also a significant difference between the 2 ports of the microinfusion manifold, but this was less pronounced. Both ports of the microinfusion manifold yielded delivery dynamics that were similar to the most downstream port of the 4-stopcock manifold. There was good correlation between dynamic data and dead volume for each of the manifolds. CONCLUSIONS: Using a traditional stopcock manifold, port selection significantly affects drug delivery dynamics for continuous infusions. The findings provide quantitative support for the concept that the most critical infusion should join the system at the manifold port closest to the patient. Port selection was less important for the microinfusion manifold and dynamics were faster compared with the second and fourth ports of the stopcock manifold. The smaller dead volumes of the microinfusion manifold minimize unwanted delays in drug delivery onset and offset allowing more precise control over drug delivery by continuous infusion.
Author information
Author/s: Moss, David R (DR); Bartels, Karsten (K); Peterfreund, Gregory L (GL); Lovich, Mark A (MA); Sims, Nathaniel M (NM); Peterfreund, Robert A (RA);
Affiliation: Department of Anesthesia and Critical Care, Massachusetts General Hospital, Boston, Massachusetts 02114, USA.
Journal and publication information
Publication Type: Comparative Study; Journal Article; Research Support, Non-U.S. Gov't
Journal: Anesthesia and analgesia (Anesth Analg), published in United States. (Language: eng)
Reference: 2009-Nov; vol 109 (issue 5) : pp 1524-9
Dates: Created 2009/10/21; Completed 2009/11/05;
PMID: 19843791, status: MEDLINE (last retrieval date: 11/5/2009, IMS Date: )
Sourced from the National Library of Medicine. Abstract text and other information may be subject to copyright.
External Links for this article
(including full text providers, if available):
Click Electronic Full-text Provider Links to see options for finding the electronic full text links to this article. Note there may be a subscription or fee required for access to the full text. See our FAQ for information on finding FREE full text articles.
This article may also be located in paper journal collections available in many libraries. Use the Journal and Publication Information above to find the full article.
MeSH headings (categories)
This article was linked to the MESH Headings shown below.
Related articles
These are the highest related articles currently in the database:
- A new bladder allows kinetic venous augmentation with a roller pump.
30 Oct 1999 - Evaluation of the FASTSTART mode for reducing start-up delay in syringe pump infusion systems.
19 Apr 2001 - An improved technique for the placement of hemodialysis catheters in the internal jugular vein.
27 Feb 1995 - Technology and clinical application of large-bore and implantable catheters.
30 Mar 1994 - Evaluation of long-term central venous patency in children with chronic venous catheters using image-directed Doppler ultrasonography.
30 May 1994 - [Utilization in chronic hemodialysis of the "implantable site" technique]
30 Dec 1993 - [Effect of catheter type on the functional survival of jugular access ports in chronic hemodialysis]
30 Dec 1993 - [Tunneled double femoral catheters: emergency vascular access for hemodialysis]
30 Dec 1993 - [Effectiveness and complications of percutaneous central venous catheters in neonatal intensive care]
30 Dec 1993 - Central venous catheter with subcutaneous injection port (Port-A-Cath): 8 years clinical follow up with children.
29 Jun 1993
Related Article Map
Legend:
- FREE Full text Article.
- Abstract only.
- Title only. More help.
See a large map of 100+ related articles.