| Research article summary (published 30 Jan 1996): |
Determination of aortic valve area in valvular aortic stenosis by direct measurement using intracardiac echocardiography: a comparison with the Gorlin and continuity equations.
Full Abstract
OBJECTIVES. This study sought to 1) show that intracardiac echocardiography can allow direct measurement of the aortic valve area, and 2) compare the directly measured aortic valve area from intracardiac echocardiography with the calculated aortic valve area from the Gorlin and continuity equations. BACKGROUND. Intracardiac echocardiography has been used in the descriptive evaluation of the aortic valve; however, direct measurement of the aortic valve area using this technique in a clinical setting has not been documented. Despite their theoretical and practical limitations, the Gorlin and continuity equations remain the current standard methods for determining the aortic valve orifice area. METHODS. Seventeen patients underwent intracardiac echocardiography for direct measurement of the aortic valve area, including four patients studied both before and after valvuloplasty, for a total of 21 studies. Immediately after intracardiac echocardiography, hemodynamic data were obtained from transthoracic echocardiography and cardiac catheterization. RESULTS. Adequate intracardiac echocardiographic images were obtained in 17 (81%) of 21 studies. The average aortic valve area (mean +/- SD) determined by intracardiac echocardiography for the 13 studies in the Gorlin analysis group was 0.59 +/- 0.18 cm2 (range 0.37 to 1.01), and the average aortic valve area determined by the Gorlin equation was 0.62 +/- 0.18 cm2 (range 0.31 to 0.88). The average aortic valve area determined by intracardiac echocardiography for the 17 studies in the continuity analysis group was 0.66 +/- 0.23 cm2 (range 0.37 to 1.01), and that for the continuity equation was 0.62 +/- 0.22 cm2 (range 0.34 to 1.06). There was a significant correlation between the aortic valve area determined by intracardiac echocardiography and the aortic valve area calculated by the Gorlin (r = 0.78, p = 0.002) and continuity equations (r = 0.82, p < 0.0001). CONCLUSIONS. In the clinical setting, intracardiac echocardiography can directly measure the aortic valve area with an accuracy similar to the invasive and noninvasive methods currently used. This study demonstrates a new, quantitative use for intracardiac echocardiographic imaging with many potential clinical applications.
Author information
Author/s: Foster, G P (GP); Weissman, N J (NJ); Picard, M H (MH); Fitzpatrick, P J (PJ); Shubrooks, S J (SJ); Zarich, S W (SW);
Affiliation: Cardiovascular Division, Deaconess Hospital, Boston, Massachusetts, USA.
Journal and publication information
Publication Type: Comparative Study; Journal Article
Journal: Journal of the American College of Cardiology (J Am Coll Cardiol), published in UNITED STATES. (Language: eng)
Reference: 1996-Feb; vol 27 (issue 2) : pp 392-8
Dates: Created 1996/02/26; Completed 1996/02/26; Revised 2007/11/15;
PMID: 8557911, status: MEDLINE (last retrieved date: 2/18/2009)
Sourced from the National Library of Medicine. Abstract text and other information may be subject to copyright.
External Links for this article
(including full text providers, if available):
Click Electronic Full-text Provider Links to see options for finding the electronic full text links to this article. Note there may be a subscription or fee required for access to the full text. See our FAQ for information on finding FREE full text articles.
This article may also be located in paper journal collections available in many libraries. Use the Journal and Publication Information above to find the full article.
MeSH headings (categories)
This article was linked to the MeSH Headings (categories) shown below.
Note: Bold headings indicate primary MeSH headings or qualifiers.
Related articles
These are the most related articles currently in our database:
- [Aortic stenosis]
13 Dec 2006 - Two-dimensional transesophageal echocardiographic determination of aortic valve area in adults with aortic stenosis.
30 Oct 1991 - Morphology and dynamic change of discrete subaortic stenosis can be imaged and quantified with three-dimensional transesophageal echocardiography.
30 Aug 1997 - Clinical evaluation of high-frequency (ultrasonic) mechanical débridement in the surgical treatment of calcific aortic stenosis.
29 Nov 1990 - Doppler echocardiographic methods to estimate severity of aortic stenosis.
13 Sep 1995 - Valve area and cardiac output in aortic stenosis: quantification by magnetic resonance velocity mapping.
30 Oct 1993 - Flow dependence of valve area in aortic stenosis: relation to valve morphology.
27 Feb 1998 - Aortic valve area discrepancy by Gorlin equation and Doppler echocardiography continuity equation: relationship to flow in patients with valvular aortic stenosis.
30 Jul 2000 - [Noninvasive quantification and classification of the severity of aortic stenosis using Doppler echocardiography]
30 Oct 1992
Related Article Map